Laboratory Turnaround Time: Split Queue, Test and Verification Time Before Buying More Equipment

Laboratory Turnaround Time: Split Queue, Test and Verification Time Before Buying More Equipment

Before you approve another analyzer, split your laboratory turnaround time into segments: receipt-to-queue, queue-to-test, test duration, repeat work, and verification-to-release. Measure waiting separately from active testing before deciding whether equipment or workflow changes would help.

This guide shows you how to build a timestamp map from specimen receipt through result release, how to stratify lab TAT by shift, test family and priority, and how to tell whether your delay comes from specimen queue, batching, analyzer utilization, repeat testing or result verification. You will also find a stratification plan, a worked illustrative example, and four Air Academy Associates resources that support the work.

Key Takeaways

  • Define turnaround time endpoints and targets for each test.
  • Separate preparation, queue, analysis, repeat work and verification time.
  • Validate timestamps before comparing shifts, priorities and test families.
  • Calculate total TAT per specimen before summarizing the results.
  • Match workflow or equipment changes to the measured constraint.

Start with the decision in front of you, because that decision shapes what data you need.

Should You Buy More Equipment or Fix Laboratory Turnaround Time First?

Should You Buy More Equipment or Fix Laboratory Turnaround Time First?

Measure demand, downtime and workflow constraints before you request capital. Assess how added capacity would address the measured delay, including queues caused by instrument congestion. A second analyzer will not directly resolve a courier scheduling delay.

Ask three questions before the purchase request leaves your desk.

  • What is actual analyzer utilization during the peak hour? Not daily average. Peak hour, by instrument.
  • How much unplanned downtime occurred last quarter? Calibration failures, QC reruns and maintenance often consume capacity you already own.
  • Which stages contribute most to long turnaround times? Investigate preparation, queue and verification delays before assuming analytical capacity is the problem.

The WHO guidance on turnaround time recommends setting maximum TATs for each examination in discussion with laboratory clients and examining the time needed for individual process steps. Use test-specific targets rather than relying only on a blended figure.

Once you accept that TAT is a sum, the next step is naming each part clearly.

How to Decompose Lab TAT From Receipt Through Result Release

How to Decompose Lab TAT From Receipt Through Result Release

Break the journey into segments that each have a start timestamp and a stop timestamp. Every segment must be owned by a specific step and a specific team. Vague segments produce vague improvement plans.

Here is a proposed starting map to adapt to your laboratory. Define whether the project measures receipt-to-release TAT or includes collection and transport; use separate records for individual tests when one specimen supports several examinations.

  1. Collection to receipt. Transport, courier scheduling and ward handling live here. Check whether collection and receipt events are available in the laboratory information system (LIS).
  2. Receipt to accessioning. The specimen is physically present but not yet in the system. Investigate registration steps and staffing coverage.
  3. Accessioning to specimen queue entry. Centrifugation, aliquoting, sorting and racking happen in this window.
  4. Queue to analysis start. Measure waiting after preparation is complete, including any delay between loading and actual analysis.
  5. Analysis start to raw result. Measure active analytical processing separately from loading and waiting.
  6. Repeat and reflex work. Flag reruns, dilutions, reflex tests and add-ons separately; these are not all avoidable defects. Record their additional time without double-counting intervals already included in analysis or review.
  7. Final result available to release. Include review waiting, verification and transmission, with separate timestamps where available.

A laboratory turnaround time dashboard study measured testing and review intervals separately and identified delays involving staffing, equipment, procedures and authorisation. Its findings support investigating local causes rather than assuming that one stage dominates every laboratory.

Segments are only useful when you can timestamp them, so the map comes next.

Building a Timestamp Map for Specimen Queue and Result Verification

A timestamp map is a simple table listing every segment, its start event, its stop event, the data source and the owner. Build it before you pull an extract so missing or ambiguous timestamps become clear. Air Academy's practical value stream mapping guide can help structure the process map.

The table below is an illustrative example only. The durations represent one fictional specimen with no repeat testing; they are not medians or client results. The repeat row shows an optional event to log separately, not a measured frequency.

Segment Start Event Stop Event Data Source Illustrative Duration
Transport Collection scan Lab receipt scan Courier log + LIS 34 min
Accessioning Receipt scan Order activated LIS 9 min
Prep and queue Order activated Analysis starts LIS + middleware 41 min
Analysis Analysis starts Raw result posted Instrument log 17 min
Repeat work Rerun flag Final raw result Middleware Not included in this example
Verification Final raw result available Result released LIS audit trail 22 min

For this fictional specimen, transport, preparation/queue and verification total 97 minutes (34 + 41 + 22), but that is not all pure waiting. Receipt-to-release TAT is 89 minutes (9 + 41 + 17 + 22), while collection-to-release TAT is 123 minutes including transport.

For real data, calculate each specimen's total first, then summarize the totals; adding segment medians does not generally give the median total. NIST's explanation of medians and percentiles describes these as summaries of ordered observations.

If a timestamp is missing, consider a short manual observation study using clearly defined events. Two weeks can be a pilot period, but choose coverage based on test volumes, shifts and the decision being made.

Raw segment medians mislead when you ignore the conditions that produced them.

A Stratification Plan for Workflow Improvement and Analyzer Utilization

A Stratification Plan for Workflow Improvement and Analyzer Utilization

The same overall median can conceal differences between shifts or test groups. Compare relevant groups before interpreting the results.

Use these stratification variables as a starting set.

  • Shift and hour of day. Compare your observed arrival peaks with quieter periods.
  • Day of week. Check for weekday and post-holiday differences.
  • Priority class. Report STAT, urgent and routine separately when they have different targets; an overall figure can supplement them.
  • Test family. Compare chemistry, hematology, coagulation, microbiology and molecular testing separately where their workflows differ.
  • Collection origin. Emergency department, inpatient ward, outpatient draw station, referring clinic.
  • Analyzer and operator. Investigate differences alongside test mix, workload and staffing before attributing them to maintenance or training.
  • Repeat status. Separate first-pass specimens from those requiring rerun or dilution.

Plot each stratum as a distribution, not a single average. The 90th percentile turnaround time measure describes the slower end of the distribution: approximately 90% of observations fall at or below it, depending on the calculation method and ties.

A published laboratory workflow improvement study reported shorter in-laboratory TAT after changes that included automation, autoverification, transport upgrades and reduced batching. Because several changes were combined, the study does not establish that workflow changes alone can replace equipment investment.

For broader project examples, Air Academy's guide to Six Sigma across different industries includes laboratory turnaround time.

Before any of this analysis counts, the clocks behind it need scrutiny.

Verifying Timestamp Reliability Before Analyzing Laboratory Turnaround Time

Check that instrument clocks, LIS servers, courier devices and middleware use consistent time settings and event definitions. Treat apparent timing anomalies as data-quality issues to investigate before calculating durations.

Run these checks first.

  • Compare system clocks across LIS, middleware and each analyzer on the same morning.
  • Investigate negative durations for clock, time-zone, event-definition, data-entry or record-matching problems.
  • Confirm what the receipt scan actually represents, since some sites scan at the door and others at the bench.
  • Check whether verification timestamps capture the review moment or the batch release moment.
  • Start with a small manual audit, such as twenty specimens across relevant shifts, and expand it if discrepancies appear; this is a pilot check, not a validated sample-size rule.

Use consistent event definitions and compare recorded timestamps with observed events before relying on the data. Resolving disagreements about when a step starts or stops makes comparisons more defensible.

The following courses can help teams develop the mapping, measurement and improvement skills used in this work.

Air Academy Associates Courses That Support Laboratory Turnaround Time Work

Air Academy Associates Courses That Support Laboratory Turnaround Time Work

These four offerings line up directly with the steps described above, from mapping specimen flow to validating timestamps and sustaining gains. Each one builds a specific capability that laboratory leaders, quality specialists and program managers can apply to real TAT projects.

IPO Courses and Six Sigma Value Stream Mapping

The VSM with IPO and SIPOC course covers process mapping, value-added and non-value-added activity, and cycle efficiency. For a laboratory project, use those tools to map specimen flow and identify handoffs and queue locations before collecting detailed data.

Statistical Process Control Short Course

The Statistical Process Control short course introduces control charts, chart selection and signals of process change. For a TAT project, choose charts suited to the data and sampling plan rather than treating every long turnaround as a special cause.

Measurement System Analysis

The Measurement System Analysis short course covers measurement-system evaluation, including repeatability, reproducibility and bias. These concepts can inform a timestamp study, but clock synchronisation and event logging also need checks by the laboratory's IT and technical teams.

Lean Six Sigma Green Belt Online Course

The Green Belt online training course covers DMAIC through self-paced modules. A laboratory team could apply the methods to a locally approved queue, repeat-work or verification project.

Course completion alone is not certification; the published certification process also includes an assessment and project approval. Confirm the current requirements with Air Academy Associates before enrolling.

A short comparison helps clarify which lever matches which finding.

Matching the Delay Type to the Right Laboratory Turnaround Time Intervention

Each segment responds to a different countermeasure. Matching them correctly saves budget and credibility.

Dominant Delay Possible Cause to Investigate Potential Intervention to Evaluate
Transport Courier schedule mismatch Resynchronize pickup times to ward draw rounds
Specimen queue Batching rules Evaluate smaller batches or continuous feed where compatible with the assay and quality requirements
Analysis True capacity limit at peak Assess peak demand, effective capacity, downtime and resilience needs
Repeat work Pre-analytical quality, dilution rules Investigate rerun triggers while retaining clinically necessary testing
Verification Manual review backlog Review staffing and evaluate validated autoverification where appropriate

For applicable disciplines, CLSI's guidance on validating autoverification systems supports a controlled design and validation process. Laboratory leadership should approve clinical workflow changes; faster release must preserve required quality checks.

Capacity remains a valid intervention when the evidence supports it, including peak-load and service-continuity needs. It need not wait until every other improvement has been exhausted.

After an intervention, remeasure the full pathway to check whether delay has shifted to another stage. Air Academy's process improvement consulting and support can help teams scope projects and assess improvement options.

Final Thoughts

Measure every segment, question every clock, then decide. When the data points to capacity, buy with confidence. Our team at Air Academy Associates can help you build that evidence through training or a consultation request.

Air Academy Associates offers Lean Six Sigma training and consulting to support laboratory process-improvement work. Contact the team to discuss the skills and support your turnaround-time project needs.

FAQs

Which Timestamps Are Needed to Analyze Laboratory Turnaround Time?

Define the TAT endpoints first, then capture collection and receipt separately where transport is in scope, accessioning, preparation complete/queue entry, analysis start and end, final result available, and release. Add review-start or transmission timestamps if those steps need separate investigation.

How Can Queue Time Be Separated From Analyzer and Verification Time?

Calculate queue time from preparation complete/queue entry to actual analysis start, test time from analysis start to result generation, and verification time from final result availability to release. Receipt-to-analysis includes accessioning and preparation as well as waiting, so do not label it pure queue time. If events are missing, use a locally agreed observation plan and avoid double-counting repeat work.

When Is More Equipment Unlikely to Solve the Delay?

More analyzers are unlikely to resolve a delay caused primarily by transport scheduling, accessioning or result review when analytical capacity is adequate. Equipment can still help with instrument congestion, downtime or automation needs, so evaluate the measured constraint and recheck total TAT after any change.

Related Articles

Posted by
Air Academy Associates
Air Academy Associates is a leader in Six Sigma training and certification. Since the beginning of Six Sigma, we’ve played a role and trained the first Black Belts from Motorola. Our proven and powerful curriculum uses a “Keep It Simple Statistically” (KISS) approach. KISS means more power, not less. We develop Lean Six Sigma methodology practitioners who can use the tools and techniques to drive improvement and rapidly deliver business results.